FLAIR, 16 September 2024
Yueyu Lin, István Fodor, Joris Meurs, Amir Khodabakhsh, and Simona M. Cristescu
Methane (CH4) emerges as an end-product of microbial fermentation catalyzed by the gut microbiota upon the incomplete digestion of polysaccharides from dietary carbohydrates. Currently, H2 and CH4 breath tests are employed to evaluate carbohydrate malabsorption conditions, such as lactose or fructose intolerance, as well as small bowel bacterial overgrowth. Here we report on the development of a custom-built breath analyzer based on a commercial OEM (original equipment manufacturer) module using tunable diode laser absorption spectroscopy (TDLAS) for real-time measurements of CH4 concentrations in exhaled breath. Since the module exhibited interferences with water and CO2, correction functions were experimentally identified, validated, and applied. Real-time CH4/CO2 profiles from human breath, as well as CH4 profile from dairy cattle breath were obtained using this system.